MODIFIED TRAPEZOIDAL AREA X VELOCITY, P3 = 15 
‘U’ parameters listed *
‘U’ parameters calculated **
U0 = channel top width 
B
U3 = area (h)
U1 = channel base width 
b
U2 = transition height 
ht
*obtain from manufacturer’s specifications.
**
calculated by OCM-3. May be viewed by accessing ‘U’ parameter.
Reference
ABSOLUTE CALCULATION,  P4 = 0
¤
 
For flows that can be calculated by the equation:
q = 1/1000 x (b + mh) x v     for h 
 d
q = 1/1000 x ((b + md)d + (B(h - d))) x v    for h 
 d
m = (B - b)/d
where :
q = flow rate, l/s
d = height of transition, cm
B = channel top width, cm
h = head, cm
b = channel base width, cm
v = flow velocity, cm/sec
RATIOMETRIC CALCULATION,  P4 = 1
¤
 
For flows that can be calculated by the equation:
q = q
cal
 x A/A
cal
 x v/v
cal
for h 
 d: A = (b + mh) x h 
A
cal
 = (b + mh
cal
) x h
cal
 
for h 
 d: A = (b + md) x d + B(h - d)
A
cal
 = (b + md) x d + B(h
cal
 - d)
m = (B - b)/d
where :
q = flow rate 
d = depth of channel
q
cal
 = flow rate at h
cal
h = head
B = channel top width, cm
v = flow velocity, cm/sec
b = channel base width, cm
¤
 Refer to Operation \ Flow Calculation.