Set Up Level 1 of the State Trigger Specification
In the next two exercises, you will set up the sequence of events that you
want the logic analyzer to follow when triggering and storing data. In this
exercise you will set Level 1 to store all data occurring before the trigger
pattern is found.
want the logic analyzer to follow when triggering and storing data. In this
exercise you will set Level 1 to store all data occurring before the trigger
pattern is found.
1
Highlight the
1
field in the State Sequence Levels box of the Trigger
menu. Press the
Select
key.
2
Highlight the field to the right of
Trigger on
, and press the
Select
key.
Highlight
b
in the pop-up menu, and press the
Done
key.
3
Press the
Done
key to close the Sequence Level 1 pop-up menu.
Setting Up Level 1 of the State Trigger Specification
Result
The state analyzer stores all samples, starting from the time you press run,
and continuing until the trigger pattern 00 is found.
and continuing until the trigger pattern 00 is found.
The b term holds
the value 00.
the value 00.
Introduction to State Analysis
Set Up Level 1 of the State Trigger Specification