HP (Hewlett-Packard) 70 Manuel D’Utilisation

Page de 140
Clear the Timing Analyzer Trigger
Clearing the timing analyzer trigger sets the specification to "don’t care".  By
doing this, the timing analyzer will trigger as soon as it is armed by the state
analyzer.
1
Press the 
Trigger
 MENU key.
2
Highlight the 
Modify Trigger
 field on the right side of the screen, and
press the 
Select
 key.
3
Highlight the 
Clear Trigger
 field, press the 
Select
 key, and choose 
All
from the pop-up menu.
Result
The timing analyzer trigger specification is set to a "don’t care" state. 
Set Up the State Trigger Specification
1
Press the 
Trigger
 MENU key.  Highlight 
Trigger MACHINE 1
 in the pop-up
menu, and press the 
Select
 key.  The Trigger MACHINE 1 menu is
now displayed.  If this is not the case, press the 
Trigger
 MENU key
again, then choose 
Trigger MACHINE 1
 from the pop-up menu.
2
Highlight the 
1
 field in the State Sequence Levels box of the Trigger
menu.  Press the 
Select
 key.
3
Highlight the field to the right of 
While storing
, and press the 
Select
 key.
Highlight 
anystate
 in the pop-up menu, and press the 
Select
 key.
4
Press the 
Done
 key to close the Sequence Level 1 pop-up menu.
Result
The state analyzer trigger specification is now:
Level 1 - While storing "anystate", TRIGGER on "a" 1 time
Level 2 - Store "anystate"
The trigger term "a" is assigned the hex value FF.
Mixed Mode - Correlation of State and Timing Data
Clear the Timing Analyzer Trigger
5-7