Agilent Technologies E4445A Manuel D’Utilisation

Page de 664
578
Chapter 8
Using the STATus System
STATus Subsystem
Us
in
g
the
ST
A
T
us
Syst
em
Condition register will set the corresponding bit in the Questionable
Frequency Event register when the condition register bit has a negative
transition (1 to 0). The variable <number> is the sum of the decimal
values of the bits that you want to enable.
Key Type:
There is no equivalent front panel key.
Factory Preset: 0
Range:
0 to 32767
Questionable Frequency Positive Transition
:STATus:QUEStionable:FREQuency:PTRansition <number>
:STATus:QUEStionable:FREQuency:PTRansition?
This command determines which bits in the Questionable Frequency
Condition register will set the corresponding bit in the Questionable
Frequency Event register when the condition register bit has a positive
transition (0 to 1). The variable <number> is the sum of the decimal
values of the bits that you want to enable.
Key Type:
There is no equivalent front panel key.
Factory Preset: 32767 (all 1’s)
Range:
0 to 32767
Questionable Integrity Register
Questionable Integrity Condition
:STATus:QUEStionable:INTegrity:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Integrity Condition register.
NOTE
The data in this register is continuously updated and reflects the
current conditions.
Key Type:
There is no equivalent front panel key.
Questionable Integrity Enable
:STATus:QUEStionable:INTegrity:ENABle <number>
:STATus:QUEStionable:INTegrity:ENABle?
This command determines which bits in the Questionable Integrity