PERFORMANCE VERIFICAITONS
  
 
Perform the following analysis, if the meter conforms to the limits listed in Table 1 the meter is functioning 
correctly.  If the meter does not conform to any of the listed limits the calibration procedure must be 
performed. 
 
Performance Verification Preparation
 
 
1.   Turn on the Calibrator, allow calibrator to warm up. Temperature Stabilization 
      should be reached after 30 minutes. 
2.   Remove battery cover and using a calibrated meter to ensure the battery measures a minimum of 7.5 V 
      DC. If the battery measures under 7.5V DC, replace the battery before beginning the performance test. 
3.    Input the values listed in Table 1 for the Standard 61-362 
 
Table 2 Performance Verification 61-362
 
Function Setting /Range 
Input 
Low Limit  
High Limit 
ACV 3.2V 
3.000V @ 50Hz 
2.936 
3.064 
ACV 3.2V 
3.000V @ 500Hz 
2.936 
3.064 
ACV 32 
30.00 @ 50Hz 
29.36 
30.64 
ACV 32 
30.00 @ 500Hz 
29.36 
30.64 
ACV 320 
300V AC @ 50Hz 
293.6 
306.4 
ACV 320 
300V AC @ 500Hz 
293.6 
306.4 
ACV 600 
500V AC @ 50Hz 
486 
514 
ACV 600 
500V AC@ 60Hz 
486 
514 
DCV 320mV  
300mV DC 
296.3 
303.7 
DCV 3.2V 
3.000V DC 
2.975 
3.025 
DCV 32V 
30.00V DC 
29.63 
30.37 
DCV 320 
300.0V DC 
296.3 
303.7 
DCV 600 
500V DC 
493 
507 
A DC 320µA 
300µA 293.9 
306.1 
A DC 3200 µA 
3000µA 2939 
3061 
A DC 32mA 
30mA 
29.39 
30.61 
A DC 320mA 
300mA 293.9 
306.1 
A DC 10 
9.00 DCA 
8.70 
9.30 
A AC 320µA 
300µA @ 50Hz 
292.1 
307.9 
A AC 320µA 
300µA @ 500Hz 
292.1 
307.9 
A AC 3200µA 
3000µA @ 50Hz 
2921 
3079 
A AC 320oµA 
3000µA @ 500Hz 
2921 
3079 
A AC 32mA 
30mA @ 50Hz 
29.21 
30.79 
A AC 32mA 
30mA @ 500Hz 
29.21 
30.79 
A AC 320mA 
300mA @ 50Hz 
292.1 
307.9 
A AC 320mA 
300mA @500Hz 
292.1 
307.9 
A AC 10A 
9A @ 50Hz 
8.64 
9.36 
A AC 10A 
9A @ 500Hz 
8.64 
9.36 
Ω 320 
300.0 297.7 
306.3 
Ω 2k 
3.000k 2.950 
3.048 
Ω 32k 
30.00k 29.52 
30.48 
Ω 320k 
300.0k 295.2 
304.8 
Ω 2M 
3.000M 2.922 
3.078 
Ω 32M 
30.00M 28.45 
31.55 
Diode Test 
500mV 
.448 
.552 
Continuity Test 
120
Ω 
Beep off 
Continuity Test 
80
Ω 
Beep on 
 