Philips V200 Manuel D’Utilisation

Page de 328
Chapter 8
8-28
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Operating Instructions
Patient Data Screen
You can view the patient data screen (Figure 8-33) by pressing the PATIENT 
DATA button in the top bar. The format of this screen is the same in VCV, PCV, 
and NPPV. Table 8-7: “Patient Data Definitions: Range, Units & Resolution” 
summarizes the definitions, display ranges, and resolution for patient data 
parameters.
Figure 8-33: Patient Data Screen
Patient Data Definitions: Range, Units & Resolution
Parameter
Description
Display Range
Resolution
PIP
Peak Inspiratory Pressure
The maximum airway pressure 
during inspiration.
-20.0 to 130
cmH
2
O (hPa)
0.1 for -20.0 to 
99.9
1 for 100 to 130
MAP
Mean Airway Pressure
The average of airway pressure 
during one full breath cycle.
-20.0 to 120
cmH
2
O (hPa)
0.1 for -20.0 to 
99.9
1 for 100 to 130
Pe End
Pressure at End Expiration
The pressure measured at the 
end of expiration.
-20.0 to 99.9
cmH
2
O (hPa)
0.1
Pi End
Pressure at End Inspiration
The pressure measured at the 
end of inspiration (or at the end 
of Insp Hold if it is >0).
-20.0 to 130
cmH
2
O (hPa)
0.1 for -20.0 to 
99.9
1 for 100 to 130
Tidal Vol.
Exhaled Tidal Volume
The volume exhaled at each 
breath. Compliance 
compensated (if enabled).
0 to 9999 ml
(BTPS)
1
Table 8-7: Patient Data Definitions: Range, Units & Resolution (Sheet 1 of 2)