Philips V200 Manuel D’Utilisation

Page de 328
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
12-1
Chapter 12. Technical Specifications
Breath Types
Modes
Volume Ventilation 
Settings, Ranges and 
Resolution
Resolution is one unit unless otherwise noted (cmH
2
O is considered 
numerically equivalent to hPa).
Breath Types
Volume Controlled Ventilation
VCV
Pressure Controlled Ventilation
PCV
Non-Invasive Positive Pressure Ventilation
NPPV
Apnea Ventilation
Table 12-1: Breath Types
Modes
Assist/Control (A/C)
VCV, PCV
SIMV
VCV,PCV
CPAP
VCV, PCV
Spont/T
NPPV
Spont
NPPV
Table 12-2: Modes
Volume Ventilation Settings, Ranges & Resolution
Setting
Range
Respiratory Rate 
1 to 80 Bpm
Tidal Volume 
50 to 2500 mL
Peak Inspiratory Flow 
3 to 140 Lpm (Compliance compensated, actual to 200 Lpm)
PEEP 
0 to 35 cmH
2
O (hPa)
PSV Pressure 
0 to 100 cmH
2
O (hPa)
Table 12-3: Volume Ventilation Settings, Ranges and Resolution (Sheet 1 of 2)