Chapter 12
12-4
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Technical Specifications
Value Entry Message
If you enter a value that is outside the operational limits of the ventilator, you 
will be shown a message window on the screen. Once you acknowledge the 
message, the ventilator returns to the value modification display without 
changing the value.
Assist/Control Ventilation
Value Entry Reasonability Checks
Item
Limit Exceeded
Message
Apnea Rate
I:E Ratio>3:1
I:E Ratio must be less than or equal to 3:1.
Check V
t
, Peak Flow, or Insp. Hold
Peak Flow
I:E Ratio>3:1
I:E Ratio must be less than or equal to 3:1.
Check V
t
, Apnea Rate, or Insp. Hold.
Tidal Volume
I:E Ratio>3:1
I:E Ratio must be less than or equal to 3:1.
Check Flow, Apnea Rate, or Insp. Hold.
Insp. Hold
I:E Ratio>3:1
I:E Ratio must be less than or equal to 3:1.
Check V
t
, Apnea Rate or Peak Flow.
Waveform
I:E Ratio>3:1
I:E Ratio must be less than or equal to 3:1.
Check V
t
, Peak Flow or Apnea Rate.
I-Time
I-Time>9 sec.
I-Time must be less than or equal to 9 sec.
Check V
t
, Peak Flow, Insp. Hold or Waveform.
Table 12-6: Assist/Control Ventilation Value Entry Reasonability Checks
Pressure Control Ventilation
Value Entry Reasonability Checks
Item
Limit Exceeded
Message
Changed
Apnea Rate
I:E Ratio>4:1
I:E Ratio must be less than 
or equal to 4:1.Check I-
Time.
Readjust the Apnea Rate or 
adjust I-Time or Rate.
I-Time
I:E Ratio>4:1
I:E Ratio must be less than 
or equal to 4:1.Check 
Apnea Rate.
Readjust the I-Time or 
adjust Apnea Rate or Rate.
I-Time
I-Time<Rise Time
I-Time must greater than or 
equal to Rise Time.
Increase the Rise Time or 
increase the I-Time.
Rise Time
Rise Time>I Time
Rise Time must be less 
than I-Time.
Decrease the Rise Time or 
increase the I-Time.
Table 12-7: Pressure Control Ventilation Value Entry Reasonability Checks