Chapter 12
Technical Specifications
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
12-9
Connectors
Alarm
Settings
Settings
High Inspiratory
Pressure
Pressure
10 to 105 cmH
2
O (hPa)
Low Inspiratory
Pressure
Pressure
3 to 105 cmH
2
O (hPa)
Low PEEP
Pressure
Pressure
0 to 35 cmH
2
O (hPa)
High Respiratory
Rate
Rate
0 to 150 Bpm
Low Exhaled
Mandatory Tidal
Volume
Mandatory Tidal
Volume
0 mL to 2500 mL
Low Exhaled
Spontaneous Tidal
Volume
Spontaneous Tidal
Volume
0 mL to 2500 mL
Low Exhaled
Minute Volume
Minute Volume
0 to 60 Lpm
High Exhaled
Minute Volume
Minute Volume
0 to 60 Lpm
High Leak (NPPV
mode only)
mode only)
0 to 60 Lpm
Alarm
Status
Indicators
Status
Indicators
Normal Indicator
High Urgency Alarm (flashing red)
Medium/Low Urgency Alarm (flashing yellow/steady yellow)
Alarm Silence
Safety Valve Open
Ventilator Inoperative (steady red)
Screen Locked
Alarms (Continued)
Table 12-17: Alarms (Sheet 2 of 2)
Connectors
Inspiratory Limb Connector
Gas outlet port: 22 mm conical male
Expiratory Limb Connector
Gas return port (on expiratory filter):
Proprietary barb fitting
Proprietary barb fitting
NOTE:
The gas return port on the ventilator is a
cylindrical port which requires mating to a
specified expiratory filter to seal the expiratory
limb.
cylindrical port which requires mating to a
specified expiratory filter to seal the expiratory
limb.
Oxygen Inlet
DISS male, DISS female, NIST, Air Liquide, or SIS Fitting
(depending on country and configuration).
(depending on country and configuration).
Table 12-18: Connectors