Chapter 13
13-32
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
NOTE:
If "On/Off Preselect" is pressed a second time, the preselection is 
switched off. 
NOTE:
If there are already three devices preselected, one of the preselected 
devices must be turned "off" before adding a new one.
7. Press "Store Preselect", then "Confirm".
8. Attach the adhesive label containing "Open Device" to the VueLink 
module. The adhesive labels MUST be placed on the module in the 
same order as they appear in the Preselection Task Window.
9. When the above steps have been completed, change the CMS to 
Monitoring mode (see "Changing Modes"). 
For more information about the VueLink module, refer to the VueLink Module 
Handbook.
Information Sent to the VueLink System
Once the CMS detects that the ventilator is connected to the module, the CMS 
automatically displays ventilator parameters. The active type of ventilation 
determines what data the CMS displays. Table Figure 13-24 defines CMS 
displays and equivalent Respironics V200 Ventilator information.
Available 
on 
CareNet 
CMS 
message
V200 
equivalent
Comments
Volume control ventilation (VCV) parameters
Mode and type of ventilation will always be displayed.
X
AWF
Flow Delivered 
flow.
X
AWP
Pressure Delivered 
pressure.
AWV
Volume
Delivered volume.
sPSV
set PSV
Set pressure support ventilation.
sPEEP
set PEEP
Set positive end expiratory pressure.
X
PIP
PIP
Measured peak inspiratory pressure.
MnAwP
MAP
Measured mean airway pressure.
Pplat
Pi End
Measured end inspiratory (plateau) pressure.
PEEP
Pe End
Measured positive end expiratory pressure.
SpMV
Spont VE
Measured spontaneous minute volume.
sTV
set TV
Set tidal volume.
MV
VE
Measured minute volume.
PtLeak
Pt Leak
Estimated patient leak.
X
TV
Vt
Measured tidal volume.
sFIO_2
set %O2
Set oxygen percentage (fractional inspired oxygen).