Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-39
Chapter 13
RS-232 Communications Option 2 (Com2)
The Respironics V200 Ventilator allows the transmission of data from the 
ventilator to the RS-232 communications interface. The Com2 option allows 
the ventilator to communicate with data systems The ventilator has two 
communications options, Com1 and Com2, which work exclusively of one 
another. 
RS-232 Configuration
When the Com2 option is installed, the RS-232 communications port is 
configured for all communications functions in the following manner:
Baud Rate 9,600
Data Bits 8
Parity  None
Stop  Bits  1
NOTE:
In diagnostics mode, with Com2 installed, the baud rate is 19,200.
To properly use the Com2 option, connect the communications cable to the 
9-pin serial connection on the back of the ventilator.
Figure 13-26: 9-pin serial port connection
WARNING:
It is the responsibility of the end user to validate the compatibility and use 
of information transmitted from the ventilator with the device to be 
connected to the ventilator.
WARNING:
The data provided via the Com2 option is for reference only. Decisions for 
patient care should be based on the clinician’s observations of the patient.
Serial Port Connection