Philips V200 Manuel D’Utilisation

Page de 328
Chapter 13
13-96
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Troubleshooting
Troubleshooting
NICO button does not 
appear on the Trended Data 
screen
Probable Cause: 
What to Do:
The ventilator may not have respiratory profile 
monitor interface software. 
Contact Respironics Technical Service: 
1-800-345-6443.
NICO button on the 
Trended Data screen is 
grayed out
Probable Cause:
What to Do:
The respiratory profile monitor interface option 
is loaded but not enabled on the ventilator.
Contact Respironics Technical Service: 
1-800-345-6443.
Ventilator is not capturing 
data sent from the 
respiratory profile monitor 
Probable Cause:
What to Do:
The bidirectional communication link is not 
functioning properly.
Check that the RS-232 serial communications 
cable (P/N 1018292) is being used.
Check cable connections.
Preform the steps listed on page 13-90 and 
page 13-91 for setting up the monitor with the 
RS-232 input/output for the ventilator.
Cycle the power on the monitor.
If there is no improvement, Contact 
Respironics Technical Service: 
1-800-345-6443.
The monitor displays the 
CHECK COMMUNICATION 
message
Probable Cause:
What to Do:
The bidirectional communication link is not 
functioning properly.
Check that the RS-232 serial communications 
cable (P/N 1018292) is being used.
Check cable connections.
If there is no improvement, Contact 
Respironics Technical Service: 
1-800-345-6443.
Probable Cause:
What to Do:
The ventilator may not have respiratory profile 
monitor software. 
If the NICO button on Trended Data screen is 
grayed out, the respiratory profile monitor 
interface option is loaded but not enabled on 
the ventilator. Contact Respironics Technical 
Service: 1-800-345-6443.
Any other condition
What to Do:
Contact Respironics Technical Service: 
1-800-345-6443.