Philips V200 Manuel D’Utilisation

Page de 328
Chapter 13
13-106
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Settings
Starting Speaking 
Mode
Prepare the patient and deflate the tracheostomy cuff. 
1. Press the Speaking Mode button.
Figure 13-59: Speaking Mode Button
The first message box appears, Turn Speaking Mode On? 
This message box explains the changes that will occur while the ventilator is 
operating in Speaking Mode.
Speaking Mode Ventilator Settings
Available in Modes
Pressure Control Ventilation (AC, SIMV, and CPAP)
Volume Control Ventilation (AC, SIMV, and CPAP)
Pressure Support Ventilation (SIMV and CPAP)
Apnea Ventilation
Patient Type:
Adults & Pediatrics
PEEP
Automatically reset to zero and low PEEP alarm is reset to zero
Trigger-type
Automatically will switch from flow triggering, if previously selected, to 
a pressure triggering of 1 cm H
2
O
Alarm Changes
Exhaled volume alarms automatically change to delivered volume 
alarms
Patient Data
Trended and displayed values based on delivered volumes
Mechanics
Respiratory mechanics feature not available
Speaking Mode 
Button