Chapter 6
Connecting Additional Equipment
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
6-3
Table 6-1: Remote Alarm Cable Kits
Connecting 
Humidifier
We recommend only those humidifiers that comply with ISO 8185 for use on 
the ventilator.
You should consult with the manufacturer(s) of the active humidification 
system (humidifier and patient circuit components) for their most recent 
recommendations regarding circuit configurations. Additionally, unless 
specifically contraindicated by these manufacturers’ recommendations, we 
recommend the use of a drop-down tube and water collection vial (water trap) 
placed between the expiratory filter and the expiratory limb of the patient 
circuit. This is necessary to prevent damage to the expiratory filter. Reference 
P/N 1006241 Water Collection System, or equivalent.
CAUTION:
Remote Alarm Cable Kits
System
Part Number
WARNING:
WARNING:
When using the Remote Alarm Port, be sure to fully test the Remote Alarm 
Port and cable by:
Verifying that annunciated alarms on the ventilator are also 
annunciated on the remote alarm.
Verifying that disconnecting the cable from the Remote Alarm port 
results in an alarm notification at the Remote Alarm.
Verifying that disconnecting the cable from the remote alarm 
results in an alarm notification at the Remote Alarm.
CAUTION: