Philips V200 Manuel D’Utilisation

Page de 328
Chapter 7
Operating Theory
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
7-3
Ventilation Modes 
Common to VCV and 
PCV
In Volume Control Ventilation and Pressure Control Ventilation, the operator 
can select between ventilation modes of Assist/Control, SIMV, and CPAP.
Assist Control Ventilation (A/C)
With the Assist/Control mode (refer to Figure 7-1 or Figure 7-2), a minimal rate 
and tidal volume (or inspiratory pressure) are set by the operator. The patient 
can trigger the ventilator at a more rapid rate, but the operator set tidal volume 
(or inspiratory pressure) is delivered during each breath. The ventilator delivers 
only mandatory breaths. Assisted breaths may be either pressure or flow 
triggered. If the trigger setting is adjusted so that the patient cannot trigger the 
ventilator, all breaths will be delivered by the ventilator at the operator set rate 
and tidal volume (or pressure).
Synchronized Intermittent Mandatory Ventilation (SIMV)
SIMV is a ventilation mode where the patient is allowed to breathe 
spontaneously and the machine attempts to deliver volume (VCV) or pressure 
(PCV) mandatory breaths in synchrony with the patient’s effort at the operator 
set rate and volume (or pressure). This is accomplished by a combination of 
spontaneous and mandatory windows that open and close. The type of breath 
delivered depends upon whether the event during the window is patient 
initiated, operator initiated or time initiated. This logic is illustrated in Table 7-
1: “SIMV Logic”.
SIMV Logic
Current
SIMV State
Inputs
Ventilator Response
Next
SIMV State
Mand
Window
Time Trigger 
(breath period 
timer expires)
Deliver mandatory breath using operator 
settings for mandatory breath type; Restart 
breath period
Mand
Window
Mand
Window
Patient Trigger; 
Operator Trigger
Deliver mandatory breath using operator 
settings for mandatory breath type
Spont
Window
Spont
Window
Patient Trigger
Deliver a spontaneous breath using 
operator settings for spontaneous breath 
type
Spont
Window
Spont
Window
Operator Trigger
Deliver a mandatory breath using operator 
settings for mandatory breath type
Spont
Window
Spont
Window
Time Trigger 
(breath period 
elapses)
Restart breath period timer
Mand
Window
Table 7-1: SIMV Logic