Chapter 8
8-16
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Operating Instructions
Figure 8-14: Changing a Setting
When a value is entered, as shown in Figure 8-14, the ventilator checks to 
assure that the value has been accepted by the operator, is within limits, and 
will not cause other settings to be out of limits. If the new value causes these 
limits to be exceeded, a diagnostic message will be displayed. (Refer to “Value 
Entry Message” on page 12-4 for 
more details about the diagnostic messages).
Selecting Alarm Limits
Follow these steps to adjust currently active alarm limits:
1. Push ALARM SETTINGS. The ventilator displays the alarm limits for 
the currently active ventilation mode (Figure 8-15).
2. Push the button for the alarm limit you want to adjust. The ventilator 
displays the current value for that alarm limit in a display insert, 
similar to Figure 8-14.
3. Press the bar or use the front panel knob to adjust the value.
4. Press ACCEPT to activate the new alarm limit or CANCEL to leave it 
unchanged.
Current or new value 
displayed digitally and 
graphically
Press ACCEPT to 
activate the new 
setting
Press CANCEL to 
leave the setting 
unchanged
Push this bar to 
decrease the value
Push this bar to 
increase the value