Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Operating Instructions
Figure 8-19: Selecting Waveforms (descending ramp selected)
Selecting Adult/Pediatric Buttons
You can configure ventilation for adult or pediatric patients (Figure 8-20) from 
the active ventilation type screen. Selecting adult or pediatric tailors the 
ventilator’s breath delivery algorithms to the selected adult/pediatric patient 
type. Selecting adult or pediatric does not change how the screens work, and 
does not change ventilator or alarm settings. The patient type selection 
determines flow output at various rise time settings for PCV, PSV, and IPAP. In 
addition, the “I-Time too long” alarms and time out for spontaneous breaths 
are set to 3.5 seconds for the adult setting and 2.5 seconds for the pediatric 
setting. Peak flow is limited to 100 LPM in all pressure-based breaths when 
using the pediatric setting.
Figure 8-20: Selecting Adult/Pediatric Controls (adult selected)
Descending ramp and square wave act as 
mutually exclusive select/non-select buttons.
Push this button 
to select 
descending ramp 
waveform
Push this button to 
select square 
waveform
Push this button to 
select pediatric controls
Push this button to 
select controls