Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Pressure compensation
↓↓
In the menu "Pressure Compensation" you can set up the parameters of pressure for the
current channel of an MLT analyzer or analyzer module. With the F3 key you can change
to the other channels.
Set up parameters:
♦ 
Select any line of variables with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable with the 
88 
 -key or the 
 -key.
♦ 
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the 
88 
 -key or
cancel and go back to the last value with the F2 key.
Line of variables "Compensation":
♦ 
Use manual pressure:
The pressure compensation of the current channel is based on the value of the line
"Manual pressure".
♦ 
Use sensor value:
The pressure compensation of the current channel is based on the value measured by
a pressure sensor. The corresponding value will be displayed in the line "Reference
pressure". This setup is not available, if no sensor is installed. Then, you will find the
standard pressure of 1013.0 hPa resp. 14.7 psig in that line.
♦ 
Disabled:
No pressure compensation will be done.
Measure
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Pressure Compensation --
Compensation:
Use manual pressure
Unit:
hPa
Reference pressure:
1013.0 hPa
Pressure:
1013.0 hPa
Manual pressure:
1013.0 hPa
Pressure measurement is:
Valid
Channel
[2nd menu page via F5 (More...)]
5.1
Analyzer Module Setup
5.1.14 Pressure Compensation