Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Inputs and outputs
↓↓
SIO-01 module
↓↓
From the menu "Local SIO Configuration Parameters" you can change to submenus to
control and set up several output configurations of the local SIO module.
General Configuration of a SIO board (the whole specification see its own manual):
♦ 
Analog outputs: minimum 2, maximum 8
♦ 
Serial interface (RS 232 or RS 485) to connect the analyzer module with
an external computer
♦ 
Three relay outputs
If there is a SIO module installed in the analyzer you have to put in "Enabled" in the line
"Module installed", otherwise "Disabled".
Set up parameters:
♦ 
Select the line of variables or any menu line with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable or change to the submenu with the 
88 
 -key or the 
 -key.
♦ 
Select the parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new adjustment with the 
88 
 -key or
cancel and go back to the last value with the F2 key.
Measure
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Local SIO Configuration Parameters --
Analog output setup...
Serial interface setup...
Configuration of Relay Outputs...
Module installed:
Enabled
Back...
5.1
Analyzer Module Setup
5.1.18 Inputs and Outputs – Local SIO
[2nd menu page via F5 (More...)]