Emerson I-0021 Manuel D’Utilisation

Page de 238
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 79
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Delay and average
↓↓
In the menu "Delay and Average" you can set up the
♦ 
analog output delay time of the measurement signal and
♦ 
averaging time to calculate the mean value of concentration
for one channel of an MLT analyzer or analyzer module. To set up the parameters of the
other channels, you have to change with the F3 key.
Set up parameters:
♦ 
Select any line of variables or the function line with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Select the variable or start the function with the 
88 
 -key or the 
 -key.
♦ 
If asked after function start: Confirm with the F2 key (Yes) or
cancel and go back to the menu page with the F4 key (Back...) or the 
 -key.
♦ 
Variable changing:
Select any digit with the 
 -key or the 
 -key and adjust a new value with
the 
↑↑
 -key or the 
↓↓
 -key
resp. select the whole parameter with the 
↑↑
 -key or the 
↓↓
 -key.
♦ 
Confirm the new value with the 
88 
 -key or
cancel and go back to the last value with the F2 key.
Measure
Channel
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Delay and Average --
Concentration output delay time:
1.000  s
Averaging time:
2  min
Set average to 0.0 !
Concentration average:
93.87  ppm
Delay function is:
Disabled
5.1 Analyzer Module Setup
5.1.19 Delay and Average
[2nd menu page via F5 (More...)]