Emerson I-0021 Manuel D’Utilisation

Page de 238
5 - 82
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
AK-protocol communication
↓↓
In the menu "AK-Communication" you can set up the parameters for the remote control of
the current channel via the serial interface. With the F3 key you can change to the other
channels of an MLT analyzer or analyzer module.
Set up parameters:
♦ 
Press the 
88 
 -key or the 
 -key to select the parameter.
♦ 
Select the parameter you want with the 
↑↑
 -key or the 
↓↓
 -key.
Options:
• 
Disabled:
No data transfer by remote control is possible.
• 
Full AK:
The data transfer is only possible for commands of the AK standard.
• 
Enabled:
The data transfer is possible for the commands of the AK standard 
and further commands like service commands.
♦ 
Confirm the parameter selected with the 
88 
 -key or
cancel and go back to the last setup with the F2 key.
Notes:
♦ 
Further informations about the "AK-communication protocol" you will find in an own
manual.
♦ 
To setup the data transfer parameters (e.g. baudrate) you have to change to the menu
"Serial Interface" (see section 5.1.18 p. 5-72 resp. 5.2.1 p. 89).
Measure
Channel
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- AK-Communication --
Remote control:
Enabled
Remote control set to "FullAK" means:
  - external range control is disabled
  - automatic range switch level calculation is disabled
  - t90 time does not change with ranges
  - all valves closed if Operation status = Ready
  - calibrations include SNGA/SEGA procedures
[3rd menu page via 2 times F5 (More...)]
5.1
Analyzer Module Setup
5.1.21 AK-Protocol Communication