5 - 90
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Configuration of Relay Outputs:
Press the 
88 
 -key or the 
 -key in the line "Configuration of Relay Outputs..." of the menu
"SIO - Configuration Parameters" to change to the corresponding menu:
There are three relays on any SIO board. The relay contact status is NO (normally
opened) by factory setting. With a jumper you also have the option to determine the relay
contact status as NC (normally closed). Informations about the whole specification of the
SIO board you can find in another manual.
In the menu "Relay Outputs" you have to determine the configuration of the three relay
outputs of the SIO!
Configuration setup:
1) Select the output number:
• 
Press the 
88 
 -key or the 
 -key in the line "Outputnumber" and adjust the output
number one, two or three with the 
↑↑
 -key or the 
↓↓
 -key.
• 
Confirm the number selected with the 
88 
 -key.
2) Choose the source module:
• 
Change with the 
↓↓
 -key to the line "Choose source module...".
• 
Change to the submenu "Analyzer Modules" with the 
88 
 -key or the 
 -key.
(illustration see next page)
All analyzer modules joint to the platform or to the MLT analyzer and the
control module are available.
• 
Go to the line you want with the 
↑↑
 -key or the 
↓↓
 -key. If there are more than eight
source modules available, you have to press the F5 key to go to another menu page.
• 
Select the tag of the module with the 
88 
 -key or the 
 -key:
The display will go back to the menu "Relay Outputs" automatically. The tag of the
module selected will appear in the line "Signal comes from".
Measure
Back...
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Relay Outputs --
Outputnumber:
1
Choose source module...
Choose signal...
Signal name:
Function control
Actual state:
ON
Invert signal:
Disabled
Signal comes from:
Control Module
5.2 I/O Module Controls
5.2.1 SIO Module – Configuration of relay outputs