Main Menu — System configuration and diagnostics
↓↓
Diagnostic menus
↓↓
Control module diagnostics
↓↓
In the menu "Control Unit Diagnostics" you can find software error messages of the
control module. If such messages are existing, you can act in the following way:
1) Make a note of the error messages.
2) Reset the message:
• 
Press the 
88 
 -key or the 
 -key in the line "Edit to reset".
• 
Set up "Reset" with the 
↑↑
 -key or the 
↓↓
 -key and confirm it with the 
88 
 -key:
The error messages will disappear, if their reason will not exist any longer.
• 
If the error messages will keep displayed:
3) System reset:
• 
Change to the menu "System Reset" with the F5 key (Reset...).
• 
Press the 
88 
 -key or the 
4) Check the error messages:
• 
Go again to the menu "Control Unit Diagnostics".
• 
If the error messages will still exist, contact your customer service, please.
• 
If the error messages will be disappeared, you should set up the report status again:
Set up "Report" in the line "Edit to reset".
Measure
Back...
Reset...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Control Unit Diagnostics --
Software error code (1 = no error):
1
And:
And:
And:
And:
And:
And:
Edit to reset:
Report
Last software error message:
And:
6.1 Diagnostic Menus
6.1.1 Control Module Diagnostics