Emerson I-0021 Manuel D’Utilisation

Page de 238
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
6 - 11
Main Menu — System configuration and diagnostics
↓↓
System Reset
↓↓
You can start the following functions in the menu "System Reset":
1) System Reset:
• 
The MLT analyzer or the platform will change to the initializing mode automatically,
if you press the 
88 
 -key or the 
 -key in the line "System reset !":
This reboot is analogous to the switching on of the control module.
2) Re-initialize network:
• 
Change to the line "Re-initialize network!" with the the 
↑↑
 -key or the 
↓↓
 -key.
• 
Press the 
88 
 -key or the 
 -key to start this function.
Caution:
♦ 
With the function "Re-initialize network !", you will delete all bindings between analyzer
modules and I/O modules.
♦ 
All configuration data of the SIO and DIO modules will disappear!
Measure
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- System Reset --
Are you sure ???
Re-initializing will destroy all the binds !!!
Re-initialize network !
System reset !
6.5 System Reset