Emerson I-0021 Manuel D’Utilisation

Page de 238
NGA-2000 System Calibration
Supplement - 24
NGA 2000
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
Functionality
4.1 Gas flow
The gas flow through the analyzers can be configured in any way:
1. only serial
2. only parallel
3. serial and parallel mixed
Sample 1
V1
AM1
AM2
AM3
Sample 2
V2
AM4
Sample 3
V3
AM5
V 4
V 5
V 6
Calibration gas mixtures
Connections for program logic and flow configurations:
• 
To each AM (channel) must be assigned a sample gas valve.
• 
During a calibration of an AM  the assigned sample gas valve will be closed and returns to
the OPEN-state after the calibration of the AM is done.
• 
The calibration gases can only flow into an AM if sample gas valve is closed.
• 
With a opened sample gas valve it is expected to actually flow sample gas.
The state of assigned sample gas valve also decides about some AM specific states!