4.4  Calibration Procedure Status
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
4 - 37
Display
Status...
Main...
Channel
Calib...
TAG
F1
F2
F3
F4
F5
Failures:
No
Maintenance-Requests:
No
Temperature:
20.0 C
Operation:
Ready
0.0
100.0
Range: 1
37.50
ppm CH4
0.00
50.00
Measure
Status...
Channel
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Analyzer Module Calibration --
Calibration procedure status...
Start span calibration procedure !
Expected span gas:
50.00  ppm
Span gas name:
Methane
Start zero calibration procedure !
Expected zero gas:
0.00  ppm
Valves...
Valve position:
Samplegas
Time & Date:
12:50:55  June 16, 1998
Measure
Cancel !
Back...
More...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Calibration Procedure Status --
Procedure status:
Ready
Current/expected gas flow:
Samplegas
Concentration in span gas units:
37.50  ppm
Last zero calibration:
Success
Last span calibration:
Success
Last span calibration was:
Fri 05-29-1998 13:37:23
Successful zero+span calibrated ranges:
1+2+3+4
Remaining procedure time:
0  s
Last zero calibration was:
Fri 05-29-1998 13:32:06
---------------------- Results ----------------------
⇒ 
⇒ 
Change to the
menu "Analyzer
Module Calibration"
Press  
F5
⇒ 
⇒ 
Change to the
menu "Calibration
Procedure Status"
Press  
Options:
Via the F5 key you can
change to a submenu,
where you can close all
valves or set up each valve
separately with:
• 
zero gas or
• 
span gas or
• 
sample gas or
• 
test gas.
⇒ 
⇒ 
Change to the menu
page "Calibration
Deviations"
Press  
F5