Emerson I-0021 Manuel D’Utilisation

Page de 238
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
5 - 5
Main Menu — Analyzer and I/O-module expert configuration
↓↓
Analyzer module setup
↓↓
Calibration parameters
↓↓
From the menu "Calibration Parameters" you can change to several submenus to set up
the zero and span gas calibration parameters and to start the different calibration methods.
Change to the submenus:
♦ 
Select the menu line you want with the 
↓↓
 -key or the 
↑↑
 -key.
♦ 
Change to the corresponding submenu with the 
88 
 -key or the 
 -key.
Alternatives:
♦ 
F1: Change to the single component display !
♦ 
F3: Change to the other available channels of an MLT analyzer or
analyzer module !
♦ 
F4: Go back to the last menu page !
Measure
Channel
Back...
TAG
F1
F2
F3
F4
F5
37.50 ppm
-- Calibration Parameters --
Span gases...
Calibration procedure setup...
Time controlled calibration...
Calibration...
Advanced calibration methods...
Tolerances...
Zero gases...
5.1 Analyzer Module Setup
5.1.1 Calibration Parameters