CALCulate Subsystem
TVS600 & TVS600A Command Reference
2–75
CALCulate:WMParameter:HREFerence:RELative
CALCulate:WMParameter:HREFerence:RELative?
Sets or queries the high reference (distal) level used for time and amplitude
measurements. The level can be expressed as a ratio or percent of the current
value for CALC:WMP:AMPLitude. This setting is used when you have set
CALC:WMP:RMEThod to RELative.
CALCulate<n>:WMParameter:HREFerence:RELative <hrefā>
CALCulate<n>:WMParameter:HREFerence:RELative?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<href> (Set relative value of HREF)
Query response
<NRf >
0.0 
 N  1.0
1
MINimum
MAXimum
<NR2>
0.0
1.0
1
The units must be specified as PCT when setting HREFerence:RELative as a percent
of the value AMPL.
0.9
Execution Error –222, “Data out of range”
Attempted to set the high reference to an illegal value.
None
Command:
CALC1:WMP:HREF:REL 95 PCT
Query:
CALC1:WMP:HREF:REL?
Response:
0.95
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples