Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
CALCulate Subsystem
2–80
TVS600 & TVS600A Command Reference
Command:
CALC1:WMP:MREF:HYST 10 PCT
Query:
CALC1:WMP:MREF:HYST?
Response:
0.1
CALCulate:WMParameter:MREFerence
CALCulate:WMParameter:MREFerence:RELative
CALCulate:WMParameter:MREFerence:RELative?
Sets or queries the middle reference (mesial) level used for time and amplitude
measurements. The level can be expressed as a ratio or percent of the current
value for AMPLitude. This setting is used only when CALC:WMP:RMEThod is
set to RELative.
CALCulate<n>:WMParameter:MREFerence:RELative <mrefā>
CALCulate<n>:WMParameter:MREFerence:RELative?
<n>
1
Query response
1
2
3
4
Not applicable
1
The calculate block number <n> selects one of four calculate blocks. If you omit <n>,
the default is calculate block 1.
<mref> (Set relative value of MREF)
Query response
<NRf >
0.0 
 N  1.0
1
MINimum
MAXimum
<NR2>
0.0
1.0
1
The units must be specified as PCT when setting MREFerence:RELative as a percent
of the value AMPL.
0.5
Execution Error –222, “Data out of range”
Attempted to set the middle reference to an illegal value.
Examples
Related Commands
Syntax
Parameters
Reset Value
Errors and Events