Tektronix TVS600 & TVS600A Series Manuel D’Utilisation
CALibration Subsystem
TVS600 & TVS600A Command Reference
2–89
CALibration:PROBe:RESults?
TVS600A Models Only
Returns the calibration status of connected level 2 probes.
1. A value of one (1) returned by the query indicates an initialized (uncali-
brated) probe.
2. A value of zero (0) indicates a calibrated probe.
3. A value of minus one (–1) indicates a probe calibration failure.
4. A value of minus two (–2) indicates a probe calibration failed because the
probe was not connected to the calibration source.
5. A value of minus three (–3) indicates a calibration failed because an
unsupported probe or no probe was installed.
CALibration:PROBe[1]:RESults?
CALibration:PROBe2:RESults?
CALibration:PROBe3:RESults?
CALibration:PROBe4:RESults?
<calibration–status>
Query response
Not applicable
<NR1>
0
1
–1
–2
–3
0
1
–1
–2
–3
Not applicable
Execution Error –241, “Hardware missing” Attempted to query PROBe3 or
PROBe4 when the instrument is configured with two channels.
PROBe4 when the instrument is configured with two channels.
None
Command:
CAL:PROB1
Query:
CAL:PROB1?
Response:
0
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples