FUNCtion and DATA Subsystems
TVS600 & TVS600A Command Reference
2–103
Query:
FUNC?
Response:
"XTIM:VOLT 1"
FUNCtion:OFF
FUNCtion:CONCurrent
FUNCtion[:ON]:ALL
DATA?
FUNCtion[:ON]:ALL
Enables all sense functions. XTIM:VOLT n functions acquire data from the input
channel <n>. You must first set FUNCtion:CONCurrent to ON before sending
FUNCtion:ALL to enable all functions. To enable individual functions, use
FUNCtion:[:ON].
[SENSe:]FUNCtion[:ON]:ALL
None
Not applicable
Execution Error –221, “Settings conflict”
Attempted to execute this command when FUNCtion:CONCurrent is OFF.
This command determines the number of records returned by “DATA?”.
Command:
FUNC:ALL
FUNCtion:OFF:ALL
FUNCtion:CONCurrent
Related Commands
Syntax
Data Types
Reset Value
Errors and Events
Dependencies
Examples
Related Commands