FUNCtion and DATA Subsystems
TVS600 & TVS600A Command Reference
2–109
FUNCtion:STATe
FUNCtion:STATe?
FUNCtion:STATe?
Sets or queries the state of a specified sense function. It operates like the
FUNCtion[:ON] and FUNCtion:OFF commands which allow you to enable or
disable a function.
FUNCtion[:ON] and FUNCtion:OFF commands which allow you to enable or
disable a function.
Channel names and function strings may be used interchangeably for parameters.
This is not standard SCPI practice and does not provide horizontal compatibility.
This is not standard SCPI practice and does not provide horizontal compatibility.
[SENSe:]FUNCtion:STATe <function>, <boolean>
[SENSe:]FUNCtion:STATe? <function>
<function>
Query response
“XTIMe:VOLTage[:DC] 1” or CHAN1
“XTIMe:VOLTage[:DC] 2” or CHAN2
“XTIMe:VOLTage[:DC] 3” or CHAN3
“XTIMe:VOLTage[:DC] 4” or CHAN4
“XTIMe:VOLTage[:DC] 2” or CHAN2
“XTIMe:VOLTage[:DC] 3” or CHAN3
“XTIMe:VOLTage[:DC] 4” or CHAN4
Not applicable
<boolean>
Query response
<NRf >
1 or ON
0 or OFF
1 or ON
0 or OFF
<NR1>
1
0
1
0
0
Execution Error –224, “Illegal parameter value”
Attempted to to enable or disable an invalid SENSe function.
Attempted to to enable or disable an invalid SENSe function.
Execution Error –241, “Hardware missing”
Attempted to enable or disable “XTIM:VOLT 3” or “XTIM:VOLT 4” when the
instrument is configured with two channels.
Attempted to enable or disable “XTIM:VOLT 3” or “XTIM:VOLT 4” when the
instrument is configured with two channels.
This command determines the number of records returned by the query DATA?.
Command:
FUNC:STAT XTIM:VOLT 1", ON
Query:
FUNC:STAT? XTIM:VOLT 1"
Response:
1
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples