OUTPut Subsystem
TVS600 & TVS600A Command Reference
2–141
Command:
OUTP:ECLT0:SOUR BTR
Query:
OUTP:ECLT0:SOUR?
Response:
BTR
OUTPut:ECLTrg<n>
OUTPut:ECLTrg<n>:POLarity
OUTPut:PCOMpensate
OUTPut:PCOMpensate?
Sets or queries whether the probe compensation signal is output on the front-
panel connector PROBE COMPENSATION. Use the command
OUTPut:PCOMpensate:FUNCtion to select the type of compensation signal.
You can select either CLOCk at 1 kHz and 500 mV peak-to-peak amplitude or
VOLT with a 500 mV DC level.
When compensating probes, use the BNC-to-probe tip adapter recommended by
the probe manufacturer and connect the probe ground lead to signal ground.
OUTPut:PCOMpensate[:STATe] <boolean>
OUTPut:PCOMpensate[:STATe]?
<boolean>
Query response
<NRf >
1 or ON
0 or OFF
<NR1>
1
0
0
None
None
Command:
OUTP:PCOM ON
Query:
OUTP:PCOM?
Response:
1