Tektronix TVS600 & TVS600A Series Manuel D’Utilisation
OUTPut Subsystem
2–144
TVS600 & TVS600A Command Reference
OUTPut:REFerence:FUNCtion
OUTPut:REFerence:FUNCtion?
OUTPut:REFerence:FUNCtion?
Sets or queries which reference signal is output on the connector REFERENCE
OUTPUT. The reference signal is intended for use during instrument verifica-
tion. You can direct either of two signals to the REFERENCE OUTPUT: the
internal 10 MHz reference clock with a 1 V peak-to-peak amplitude or the 8.0 V
DC level. Use OUTPut:REFerence to enable the signal output. The reference
signal output is disabled when you start acquisition or reset the instrument.
OUTPUT. The reference signal is intended for use during instrument verifica-
tion. You can direct either of two signals to the REFERENCE OUTPUT: the
internal 10 MHz reference clock with a 1 V peak-to-peak amplitude or the 8.0 V
DC level. Use OUTPut:REFerence to enable the signal output. The reference
signal output is disabled when you start acquisition or reset the instrument.
OUTPut:REFerence:FUNCtion <function>
OUTPut:REFerence:FUNCtion?
<function>
Query response
CLOCk
VOLTage
VOLTage
CLOC
VOLT
VOLT
CLOC
Execution Error –141, “Invalid character data”.
Attempted to set the calibration reference function to an illegal value.
Attempted to set the calibration reference function to an illegal value.
None
Command:
OUTP:REF:FUNC VOLT
Query:
OUTP:REF:FUNC?
Response:
VOLT
OUTPut:REFerence
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands