Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
STATus Subsystem
2–170
TVS600 & TVS600A Command Reference
STATus:SESR:QENable
STATus:SESR:QENable?
Sets or returns the contents of the Event Status Enable Register (ESER). When
you set an event bit true (1) in the ESER, the corresponding event the Standard
Event Status Register generates a message in the Status Queue when the event
line changes from false to true (0 to 1).
STATus:PRESet sets the register bits to hexadecimal 3C.
STATus:SESR:QENable <mask>
STATus:SESR:QENable?
<mask>
Query response
<NRf > | <Non-decimal numeric>
 N 
 #HFFFF
None
Not applicable
Execution Error –222, “Data out of range”
Attempted to set the queue enable register to an illegal value.
None
Command:
STAT:SESR:QEN #H7C
Query:
STAT:SESR:QEN?
Response:
124
This response is the decimal equivalent of hexadecimal 7C.
*ESE
*ESR
STATus:PRESet
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands