Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
SYSTem Subsystem
2–202
TVS600 & TVS600A Command Reference
SYSTem:ERRor:CODE? (Query Only)
Returns the next event codes stored in the waveform analyzer Status Queue. The
returned error code does not include the associated event message. Errors and
events are cleared from the Status Queue at power up, upon receipt of the
command *CLS, and upon reading the last item from the event queue. To return
all error codes use SYSTem:ERRor:CODE:ALL?. Use SYSTem:ERRor? to
return an error code and its event message.
Refer to the TVS600 & TVS600A Series Waveform Analyzers User Manual for a
detailed description of the Status Queue and a complete list of system events.
SYSTem:ERRor:CODE[:NEXT]?
<code>
Query response
Not applicable
<NR1>
Not applicable
0, “No error”
Attempted to read an event when the queue is empty.
Device Specific Error –350, “Queue Overflow”
The Status Queue has overflowed or run out of memory.
None
Query:
SYST:ERR:CODE?
Response:
-221
SYSTem:ERRor?
SYSTem:ERRor:CODE:ALL?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples