Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
TEST Subsystem
TVS600 & TVS600A Command Reference
2–211
TEST:RESults? (Query Only)
Returns the failure code for the last self-test command that was executed. The
returned failure code identifies the first test that failed. See the Parameters table
for the possible responses. Use the command *WAI, *OPC, or *OPC? to delay
execution of the TEST:RESults? command until the self test completes.
Completion of the tests clears the TESTing flag in the Operation Status register.
<failure_code>
Query response
Not applicable
<NR1>
–1
(test in progress)
0
(no failure)
1000 – 1999
(Self Test failure)
Not applicable
None
None
Query:
Response:
TEST
TEST:RESults:VERBose?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands