Tektronix TVS600 & TVS600A Series Manuel D’Utilisation
TEST Subsystem
TVS600 & TVS600A Command Reference
2–211
TEST:RESults? (Query Only)
Returns the failure code for the last self-test command that was executed. The
returned failure code identifies the first test that failed. See the Parameters table
for the possible responses. Use the command *WAI, *OPC, or *OPC? to delay
execution of the TEST:RESults? command until the self test completes.
Completion of the tests clears the TESTing flag in the Operation Status register.
returned failure code identifies the first test that failed. See the Parameters table
for the possible responses. Use the command *WAI, *OPC, or *OPC? to delay
execution of the TEST:RESults? command until the self test completes.
Completion of the tests clears the TESTing flag in the Operation Status register.
<failure_code>
Query response
Not applicable
<NR1>
–1
–1
(test in progress)
0
(no failure)
1000 – 1999
(Self Test failure)
Not applicable
None
None
Query:
Response:
TEST
TEST:RESults:VERBose?
TEST:RESults:VERBose?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands