TRIGger[:A] Subsystem
TVS600 & TVS600A Command Reference
2–243
Changing the trigger source from INTernal to EXTernal or from one INTernal
source to another INTernal source with a different VOLTage:RANGe setting may
change the TRIGger:LEVel setting to accommodate a new vertical range.
Command:
TRIG:SOUR INT1
Query:
TRIG:SOUR?
Response:
INT1
ARM:SOURce
TRIGger:B:SOURce
TRIGger:TYPE
TRIGger:TYPE?
Sets or queries the type of triggering to use for the next acquisition. The
available types are as follows:
 EDGE specifies the default triggering. A trigger event occurs when a signal
passes through a specified voltage level on the specified rising or falling
slope. The standard SCPI trigger commands control this mode.
 LOGic–A trigger event occurs when the specified combination of input
signals occur. (TVS600A models only)
 PULse generates a trigger event when a specified pulse is detected. It is
controlled by commands in the TRIGger[:A]:PULSe subsystem.
 SHOLdtime–A trigger event occurs when the specified clock-to-data setup
and hold time is violated. (TVS600A models only)
 TRANsition–A trigger event occurs when the input signal fails to transition
between two thresholds in the specified time. (TVS600A models only)
TRIGger[:A]:TYPE <trigger_type>
TRIGger[:A]:TYPE?
<trigger_type>
Query response
EDGE
LOGic
PULSe
SHOLdtime
TRANsition
EDGE
LOG
PULS
SHOL
TRAN
Dependencies
Examples
Related Commands
Syntax
Parameters