Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
TRIGger:LOGic Subsystem
2–266
TVS600 & TVS600A Command Reference
TRIGger:LOGic:STATe:SLOPe
TRIGger:LOGic:STATe:SLOPe?
TVS600A Models Only
Sets or queries the slope of the clock (Channel 4) for logic state triggering.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is LOGic and TRIG-
gger:LOGic:CLASs is STATe.
TRIGger[:A]:LOGic:STATe:SLOPe <slope>
TRIGger[:A]:LOGic:STATe:SLOPe?
<slope>
Query response
NEGative
POSitive
NEG
POS
None
POS
Command Error –141, “Invalid character data”
Attempted to set SLOPe to an illegal value.
None
Command:
TRIG:LOG:STAT:SLOP NEG
Query:
TRIG:LOG:STAT:SLOP?
Response:
NEG
TRIGger:TYPE
TRIGger:LOGic:CLASs
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands