TRIGger:LOGic Subsystem
2–266
TVS600 & TVS600A Command Reference
TRIGger:LOGic:STATe:SLOPe
TRIGger:LOGic:STATe:SLOPe?
TRIGger:LOGic:STATe:SLOPe?
TVS600A Models Only
Sets or queries the slope of the clock (Channel 4) for logic state triggering.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is LOGic and TRIG-
gger:LOGic:CLASs is STATe.
gger:LOGic:CLASs is STATe.
TRIGger[:A]:LOGic:STATe:SLOPe <slope>
TRIGger[:A]:LOGic:STATe:SLOPe?
<slope>
Query response
NEGative
POSitive
POSitive
NEG
POS
POS
None
POS
Command Error –141, “Invalid character data”
Attempted to set SLOPe to an illegal value.
Attempted to set SLOPe to an illegal value.
None
Command:
TRIG:LOG:STAT:SLOP NEG
Query:
TRIG:LOG:STAT:SLOP?
Response:
NEG
TRIGger:TYPE
TRIGger:LOGic:CLASs
TRIGger:LOGic:CLASs
Syntax
Parameters
Suffix Units and
Multipliers
Reset Value
Errors and Events
Dependencies
Examples
Related Commands