TRIGger:TRANsition Subsystem
2–294
TVS600 & TVS600A Command Reference
TRIGger:TRANsition:CLASs
TRIGger:TRANsition:CLASs?
TRIGger:TRANsition:CLASs?
TVS600A Models Only
Sets or queries the class of transition triggering selected. RUNT: the instrument
triggers when a pulse crosses one of two thresholds but does not cross the second
threshold before recrossing the first. SLEWrate: the instrument triggers when an
edge crosses two thresholds. All classes may be time qualified.
triggers when a pulse crosses one of two thresholds but does not cross the second
threshold before recrossing the first. SLEWrate: the instrument triggers when an
edge crosses two thresholds. All classes may be time qualified.
:A is a predefined alias for :SEQuence[1].
This command only has effect when TRIGger:TYPE is TRANsition.
TRIGger[:A]:TRANsition:CLASs <transition_class>
TRIGger[:A]:TRANsition:CLASs?
<transition_class>
Query response
RUNT
SLEWrate
SLEWrate
RUNT
SLEW
SLEW
None
RUNT
Command Error –141, “Invalid character data”
Attempted to set class to an illegal value.
Attempted to set class to an illegal value.
None
Command:
TRIG:TRAN:CLAS SLEWRATE
Query:
TRIG:TRAN:CLAS?
Response:
SLEW