VOLTage Subsystem
2–304
TVS600 & TVS600A Command Reference
VOLTage:RANGe[:UPPer]
VOLTage:RANGe[:UPPer]?
Sets or queries the most positive end of the amplifier voltage range. You should
set the amplifier voltage range to match the signal amplitude you expect to
acquire with the digitizer. The string :UPPer may be omitted as shown in the
examples that follow.
The range of permitted values for <upper> depends on the attached probe.
Attached probes modify the full-scale range by the attenuation factor of the
probe. For example, a 10X probe will cause the 100 mV range to become a 1 V
range. This in turn causes the minimum and maximum offset values to change
from 
1 V to 10 V.
[SENSe:]VOLTage<n>[:DC]:RANGe[:UPPer] <upper>
[SENSe:]VOLTage<n>[:DC]:RANGe[:UPPer]?
<n> (Channel number)
1
Query response
1
2
3
4
Not applicable
1
The input channel number <n> is used by the commands INPut<n> and VOLTage<n>.
If you omit <n>, the default is channel 1.
<upper> (Upper limit of voltage range)
1
Query response
<NRf >
<NR3>
1
The default multiplier for the <upper> parameter is V for volts. You can also use MV
for millivolts.
0.5E+0
Execution Error –222, “Data out of range”
Attempted to set upper to an illegal value.
Execution Error –241, “Hardware missing”
Attempted to program VOLTage3 or VOLTage4 on an instrument configured
with only two channels.
Syntax
Parameters
Reset Value
Errors and Events