AADVance:RECord:STARt
AADVance:RECord:STARt?
Sets or queries the number of the first auto-advance waveform record to transfer
in response to the commands DATA?, TRACe?, and TRACe:COPY?. The
maximum :STARt value depends on the number of acquired records. A setting of
zero (0) selects the last waveform record for transfer. Negative settings select
waveform records referenced from the last one, the zero record. For example, the
–1 record is the second from last and the –2 record is third from last.
Note that auto-advance acquisition always starts with record number one.
[SENSe:]AADVance:RECord:STARt <start>
[SENSe:]AADVance:RECord:STARt?
<start>
Query response
<NRf >
–MAX + 1 to –1
(Number before last 
record)
0
(Last record)
1 to MAX
(Number from first record)
MINimum
MAXimum
<NR1>
1
<depends on configuration>
1
Execution Error –222, “Data out of range”
Attempted to set start to an illegal value.
None
Command:
AADV:REC:STAR 100
Query:
AADV:REC:STAR?
Response:
100