Tektronix TVS600 & TVS600A Series Manuel D’Utilisation

Page de 410
CALCulate Subsystem
2–30
TVS600 & TVS600A Command Reference
" (NONE)
Execution Error –141, “CALC1:FEED2 FTM”
Attempted to set FEED2 to an invalid trace name.
Execution Error –224, “Illegal parameter value, invalid sense function”
Attempted to set FEED2 to an invalid sense function.
Execution Error –241, “Hardware missing”
Attempted to set feed to “XTIM:VOLT 3” or “XTIM:VOLT 4” when the
instrument is configured with two channels.
None
Command:
CALC1:FEED2 XTIM:VOLT 1"
Query:
CALC1:FEED2?
Response:
"XTIM:VOLT 1"
CALCulate:PATH
CALCulate:FEED2:CONText
CALCulate:FEED2:CONText?
Sets or queries the measurement parameter block, one of CALC1–CALC4, used
to characterize the FEED2 data.
Any of the four CALC blocks, CALC1–CALC4, may be used  interchangeably
for the context parameter <calc_block>.
CALCulate<n>:FEED2:CONText <calc_block>
CALCulate<n>:FEED2:CONText?
Reset Value
Errors and Events
Dependencies
Examples
Related Commands
Syntax