NI 5412 Specifications
4
ni.com
Frequency and Transient Response
Bandwidth
20 MHz
–3 dB
Digital 
Interpolation 
Filter
Software-selectable Finite Impulse Response (FIR) filter. 
Available interpolation factors are 2, 4, or 8.
Passband 
Flatness
Low-Gain and High-Gain Amplifiers Path
±1.0 dB from DC to 6 MHz
Pulse 
Response
Path
All values are 
typical. 
Measured with a 
1 m RG-223 
cable. 
Low-Gain Amplifier
High-Gain Amplifier
Rise/Fall Time
<20 ns
<20 ns
Aberration
<5%
<5%
Table 1.   (Continued)
Specification
Value