   17 CHECK AND ADJUSTMENT
17-12
17.5 Adjustment of Compensation Systematic Error of Instrument
(Only for GPT-2003/2005/2006)
Operating procedure
Operation
Display
1
Level the instrument properly with the
plate level.
2
While pressing the [F1] key, turn power switch ON.
[F1]
+
Power ON 
3
Press the [F3] key.
[F3]
4
Press the [F1] key.
[F1]
5
Collimate target A (around 0° in horizontal within 
±3°) in normal telescope setting (FACE (1)).
Collimate
A
(Normal)
6
Press the [F4](SET) key. *1)
The sample display shows that the measurement 
is made 5 times in FACE 1.
[F4]
7
Turn the telescope in reverse telescope setting 
(FACE (2)).
8
Collimate target A.
Turn
telescope
9
Press the [F4](SET) key.
Repeat the procedures in step 
8
 and 
9
 so that the 
count of measured times matches to the one in 
FACE(1). *2),3),4)
The title display will be shown automatically.
[F4]
10
Collimate target B (more than ±10° from the level) 
in reverse telescope setting
(FACE(2)). *5)
11
Press the [F4](SET) key. *5)
Collimate
B
(Reverse)
[F4]
12
Turn the telescope in normal telescope setting 
(FACE(1)).
13
Collimate target B.
Turn
telescope
[F4]
  
ADJUSTMENT MODE  
 F1:V ANGLE 0 POINT
 F2:INST. CONSTANT
 F3:V0 AXIS      
  
AV0 AXIS
 F1:MEASUREMENT
 F2:CONSTANT LIST
  
V0/AXIS ADJUSTMENTS
  
ERROR CORRECTION
(A)COLLIMATION
(B)H AXIS
  
FRONT FACE1       /0
 V:   89°55'50"
 LEVEL ±0
 SKIP            SET
  
REVERSE FACE2     /5
 V:  270°04'20"
 LEVEL ±0
                 SET
  
(B)HORIZONTAL AXIS
  
REVERSE FACE 2    /0
 V: 270°04'20"
 LEVEL ±10°
                 SET
  
FRONT FACE 1      /5
 V:  89°55'50"
 LEVEL ±10°
 SKIP            SET