Recalling the memory data 
Confirm that the “Data memory (data)” parameter is set to “2”. 
1  Press and hold the PRINT key until    reCall is displayed, then release the key. 
2  Press the PRINT key to enter the memory recall mode. The type of data appears in the upper 
left of the display as shown to the right: 
  Recall the data in memory using the following keys 
RE-ZERO key 
To proceed to the next data set. 
MODE key 
To go back to the previous data set. 
PRINT key 
To transmit the current data using 
the RS-232C interface. 
With SAMPLE held down, press the CAL key 
 
To delete the data currently displayed.   
Note: Deleting the data will not increase the number of data that can be stored. 
CAL key 
To exit the memory recall mode. 
3  Press the CAL key to return to the weighing mode. 
 
Transmitting all memory data at one time 
Confirm that the “Serial interface (5if)” parameters are set properly. For details, refer to “9. 
FUNCTION TABLE” and “15. CONNECTION TO PERIPHERAL EQUIPMENT”. 
1  Press and hold the PRINT key until   reCall is displayed, 
then release the key. 
2  Press the SAMPLE key to display    out . 
3  Press the PRINT key to display Sout no  with “no” blinking. 
5  Press the RE-ZERO key to display  out go  with “go” 
blinking. 
6  Press the PRINT key to transmit all data using the RS-232C 
interface. 
7  The balance displays   Clear when all data is transmitted. 
Press the CAL key to return to the weighing mode. 
 
Deleting all memory data at one time 
1  Press and hold the PRINT key until    reCall    is displayed, then release the key. 
2  Press the SAMPLE key several times to display    Clear . 
3  Press the PRINT key to display SClr no  with “no” blinking. 
4  Press the RE-ZERO key to display SClr go  with “go” blinking. 
5  Press the PRINT key to delete all data 
6 The balance displays  end , then    reCall when all data is deleted. 