Philips M1351A Manuel D’Utilisation

Page de 189
Recorder Board
Chapter 11    Replacing Parts    
145
Recorder Board  
To remove the recorder board: 
Remove the top cover. 
1. Disconnect the cables G, H, I, K, and L. 
2. Remove the three screws (1 x S1, 2 x S2). The S2 screws are longer than the S1 screw. 
3. Remove the cable J. 
4. Lift out the board. 
To refit the recorder board: 
1. Slide the board into the slot. To give a good electrical contact, the board is tensioned 
when screwed down. 
2. Refit the screws and screw down the board.
Figure 11-12Refitting the Recorder Board
After refitting the board, carry out the Parameter Test (see page 68).