Tektronix AM700 Manuel D’Utilisation

Page de 271
STATus:OPERation:SYSTem Registers
3–152
AM700 Audio Measurement Set Programmer Manual
STATus:OPERation:SYSTem Registers
STATus:OPERation
Table 3–10: STATus:OPERation 
Condition definition
Condition number
Status
Calibration
100
Set if calibration is occurring
Sweep
103
Set when a sweep is started;
cleared with a sweep is
finished
Trigger for Monitor and FFT
105
Status Bits:
3
Waiting for Mtone Ch 1
4
Waiting for Mtone Ch 2
5
Triggered Ch 1
6
Triggered Ch 2
STAT:OPER:SYST Summary
Table 3–11: System conditions 
Condition Definition
Condition Number
Summary bits
Floppy inserted
300
Set when floppy disk is
inserted; cleared when
removed
Spooling
302
Set when spooling a print
job; cleared when finished
Sound play
303
Set when the DSP is
performing the “sound” func-
tion
Remote Sense
304
Mirrors the internal contact
closure input