•  For 15 seconds the motor is de-energized and allowed to coast. 
•  For the next 2 minutes 220 VAC is sent from terminal 3 on connector X11 
through the normally closed auxiliary contacts of contactor K50 to the K5 
contactor coil  
  K5 contactor closes and the motor turns CCW 
  The normally closed auxiliary contacts on K5 open preventing the K50 
contactor from energizing. 
•  NOTE: If the door switch (S7) opens by the door opening, then the fan circuit 
is de-energized.  Then the motor will be pulsed in the opposite direction to 
“break” the fan.  This is to stop the fan quicker for the operator. 
c
  The steam generator circuit is energized.  Steam production takes precedence in 
the Rethermalization mode. 
•  220 VAC is sent from terminal 4 of connector X12 on the Control Board 
(A10) through the normally closed auxiliary contacts of K4 to the coil of 
Contactor K1 until the bypass probe (B5) senses 68 degrees F. 
  The contactor K1 closes and 220 VAC is sent to the 9.4 KW element (E1) 
  The normally closed auxiliary contact of K1 opens preventing the Hot Air 
Contactor K4 from energizing  
•  220 VAC is sent from terminal 8 of connector X11 on the Control Board 
(A10) through the normally closed auxiliary contacts of K44 to the coil of 
Contactor K2 until the bypass probe (B5) senses 68 degrees F. 
  The contactor K2 closes and 220 VAC is sent to the 6.27 KW element 
(E2) 
  The normally closed auxiliary contact of K2 opens preventing the Hot Air 
Contactor K4 from energizing 
  If the reduced power option is chosen then the K2 Contactor is not 
energized. 
d
  After the bypass probe (B5) is satisfied then the hot air heat circuit is energized to 
increase the cabinet temperature to the setting. 
•  220 VAC is sent from terminal 3 of connector X12 on the Control Board 
(A10) through the normally closed auxiliary contacts of K2 and K1 to the coil 
of Contactor K4 until the cabinet reaches the set temperature. 
  The contactor K4 closes and to legs of the 220 VAC is sent to the 18 KW 
element (E4) 
  The normally closed auxiliary contact of K4 opens preventing the Steam 
Contactor K1 from energizing  
•  220 VAC is sent from terminal 7 of connector X11 on the Control Board 
(A10) to the coil of Contactor K44 until the cabinet reaches the set 
temperature. 
  The contactor K44 closes and the third leg of 220 VAC is sent to the 18 
KW element (E4) 
  If the reduced power option is chosen then the K44 Contactor is not 
energized.