Aethra d2000 lite Manuel D’Utilisation

Page de 186
ISDN TESTER AND ANALYZER D2000
PRO
/D2000Lite/D2000                                                                                USER’S MANUAL
 
 
3.7.1.  Setting Up the U Interface Test 
Define the following parameters for a U interface test:   
 
 
Load (mW): Set the load in milliwatts inserted on the line. Use the 1100 mW default 
value when no other value is requested. 
Vmin threshold: Set the minimum voltage value under which the instrument resets 
its line interface. Use the 28 V (ETR080) default value when not other value is 
requested. 
Line Code: Set the line code at 2B1Q or 4B3T. NT-U (BRI) only 
Power supply bit PS1: Choose PS1 as indicated in the table below. Only for 2B1Q. 
Power supply bit PS2: Choose PS2 as indicated the table below. Only for 2B1Q. 
PS1 
PS2 
Description 
Primary power and back-up battery failure. 
Primary power failure. 
Back-up battery failure. 
Normal primary power and back-up battery.  
Table 3-2 PS Bit Assignments 
3.7.2.  Running a U Interface Test 
When you have defined the setup for the U-Interface test, continue as follows: 
 
Select ‘Use this setup for testing’. 
64
 